Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Animals (Basel) ; 11(9)2021 Aug 27.
Artigo em Inglês | MEDLINE | ID: mdl-34573484

RESUMO

Serum Amyloid A (SAA) is one of the most sensitive tests to detect inflammation in cats. In this study, two point-of-care assays for SAA measurements in cats (FUJI DRI-CHEM IMMUNO AU CARTRIDGE vf-SAA (method A), and CUBE-VET analyser (Method B), were analytically evaluated. Regarding the imprecision precision only the method A showed intra-assay and inter-assay CV < 10% at all concentrations. Both assays showed linearity with r close to 1 and the recovery were in the range of 81-112% for assay A and 85-125% for assay B and the limit of detection were 3.75 and 0.5 mg/dL for method A and B, respectively. A previously validated method for SAA quantification SAATIA; LZ-SAA (method C) was used as gold-standard to evaluate the accuracy of the assays. Significant correlations (p < 0.0001) were found between assays A and C (r = 0.94) and B and C (r = 0.91). In addition, an overlap performance test was made using serum samples from cats with non-inflammatory and cats with inflammatory. Both assays showed higher median SAA concentrations in cats with inflammatory diseases than in cats without inflammatory diseases (p < 0.0001). In conclusion, this manuscript provides data about the possible application of two point-of-care assays for the measurement of SAA concentration in cats.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...